5. Computer Remote Control
5-4 UniSite-xpi User Manual
Suspending CRC Mode CRC Mode can be suspended temporarily to allow you to go into
terminal mode to view or change parameter settings or the device data in
memory. If the User Menu Port is set to T (Terminal port) press
S
HIFT +Z + ENTER to suspend CRC Mode. If you are controlling the
programmer from the Remote port, send the 49] command.
Halting CRC
Operations
To halt any command or any ongoing CRC operation, use one of the
following commands from the Remote port. Neither of the following two
commands requires an E
NTER. Both commands are immediate and both
terminate any preceding command operation.
ASCII
Command Hex Code Description
E
SC 1B Causes the programmer to unconditionally
halt any operation except a binary transfer.
B
REAK n/a Causes the programmer to unconditionally
halt any operation in progress. This includes
all data communications transfers. The data
line must be held in the spacing condition for
110 ms to 700 ms.