EasyManua.ls Logo

DencoHappel MATRIX OP71 - Page 76

Default Icon
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating MATRIX control panel OP71 MATRIX Control Panel OP71
76 PR-2009-0021-GB • Subject to modifications • R5-04/2016
Sensor inputs Use
The following intended uses for the sensor input can be selected in the "Use" submenu:
If one of the following fan-speed control systems is selected in the "Type of control"
menu (see page 70), the use of the input for supply air or extract air must be deac-
tivated:
Fan speed setting with control panel
Temperature controlled
Without supply-air fan/without extract-air fan
Supply-air fan follows extract-air fan/extract-air fan follows supply-air fan
Sensor inputs Input type
If the following points are selected in the "Use" submenu:
Measuring duct pressure
Measuring volume flow
Volume flow measurement with pressure sensor
the suitable voltage range for the used sensor can be selected in the submenu "input
type". Factory setting is 0-10V. The submenu "Input type" is not shown for the other
uses.
Sensor inputs Range start/range end
If a sensor input is activated in the "Use" submenu, a range start and a range end (e.g.
0-500 Pa or 30-100%) must be specified. In this case the external setting or the mea-
surable range of the sensor is displayed linearly on the selected input signal (e.g. 0-
10V).
Depending on the selected use, different physical variables can be entered, as the
following table demonstrates
of the input for
Use
Duct pressure
Use for Display:
Disabled Disabled
Speed setpoint range 0-10V Speed setpoint
Measuring duct pressure Duct pressure
Measuring volume flow Volume flow
Volume flow measurement with pressure sensor volume flow<-pressure
Input type
0..10 Volt
2..10 Volt
m³/h
m³/h
Range start
1000
Range end
20000
Application: Value in:
Speed setpoint range 0-10V Percent [%] of the
maximum speed
Measuring duct pressure Pascal [Pa]
Measuring volume flow m
3
/h
Volume flow measurement with pressure sensor Pa

Table of Contents