Operation
Calibrating
4,1 Vol%
The screen displays the test value of the CO2 concen-
tration FCO2.
This value must agree to within ±3 Vol.% with the speci-
fication on the test filter.
Example: 4.1 Vol.% on the filter:
permitted value range: 3.8 to 4.4 Vol.%
If the test value is outside the permitted tolerance, the
test gas must be checked or calibrated.
Push the CO2 sensor back on to the cuvette.
Checking CO2 calibration with calibration gas
if the specified calibration value was not met when
testing with the test filter
at least once per half-year.
Switch on Evita 4. Wait about 3 minutes for the CO2
sensor to complete its warm-up phase.
Press the »Calibration« key.
Display (example):
Touch the »CO2« screen key.
141 37 207