Medicament nebuliser (option)
1 Remove the nebuliser hose from the nebuliser and
from the port on the device.
2 Remove the medicament nebuliser from the hose
system or
2 Dismantle the medicament nebuliser from the
paediatric hose system.
3 Remove the catheter connector (ISO cone ø15 / ø11)
from the inlet.
4 Remove adapter (ISO cone ø22 / ø11) from the
outlet.
5 Remove corrugated hose from the adapter.
● Dismantle the medicament nebuliser in accordance
with its specific Instructions for Use.
● Prepare the individual parts of the medicament
nebuliser and the adapting components for
disinfecting and cleaning in the autoclave.
Ventilation hoses
● Remove from the adapters and ports.
● Remove the water traps from the ventilation hoses.
Remove the collecting jars from the water traps.
● Prepare the ventilation hoses, water traps and their
collecting jars and the Y-piece for cleaning in the
autoclave.
Flow sensor
● Tilt the control unit upward.
6 Push the flow sensor to the left as far as it will go
and
7 pull out.
The flow sensor cannot be disinfected/cleaned by
autoclaving and cannot be sterilised by the hot steam
method.
● Disinfect the flow sensor for about 1 hour in 70 %
ethanol solution.
Expose the sensor to air for at least 30 minutes.
Otherwise residual alcohol could damage the
sensor beyond control due to ignition during
calibration.
● The flow sensor may be reused as long as calibration
can be carried out successfully.
137