121
Configuration
Ventilation
Setting the start-up values for the
»Pressure, O
2, I:E« ventilation parameter
● Press »Configuration« key.
● Touch »Ventilation« screen key.
Enter access code 3032:
● Touch the corresponding screen key.
● Touch the »Start up settings« screen key.
Display (example):
● Touch the »Pressure, O2, I:E« screen key.
Starting values of »Pressure, O2, I:E«
1)
∆PASB = PASB - PEEP
The hospital-specific start-up settings can be entered in
the table.
To change the start-up values of »Pressure, O2, I:E«:
● Touch the screen key for the parameter to be
changed.
● Change value = turn rotary knob.
● Confirm value = press rotary knob.
PEEP
mbar
Pinsp
mbar
∆PASB
1)
mbar
Pmax
mbar
O2
Vol.%
I:E
Factory
settings
515040301:2
Hospital-
specific
settings
........................ ........................ ........................ ........................ ........................ ........................
181 37 207