Evita Remote
Optional remote control unit (Remote Pad)
The kit may only be installed and programmed by
specialists.
For parallel, remote operation of the following LED and
key functions:
1 Red LED – to indicate warning messages
2 Yellow LED – to indicate caution and advisory
messages
3 »gg
gg
« key – to suppress the alarm tone for approx.
2 minutes
4 »Alarm Reset« key – to acknowledge alarm messages
5 » Neb.« key – to start and end medicament
nebulisation
6 »O2 suction« key – for bronchial suctioning
7 »Insp. hold« key – for sustained, manually induced
inspiration
8 »Exp. hold« key – for extended and sustained
expiration
The function of the respective LEDs and keys is the same
as that of the corresponding elements on the front panel
of Evita 4 and is described in the application chapters of
the Instructions for Use.
Connection
● Plug the lead of the Remote Pad into the socket » «
on the rear of Evita 4. The plug can be connected or
disconnected at any time without impairing operation
of Evita 4.
1
2
3
4
6
5
7
8
38
Preparation
Evita Remote