Dräger Evita 4 - Before Reusing on Patient; Maintenance Intervals

Dräger Evita 4
192 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Humidifier
Assemble in accordance with separate Instructions
for Use.
Before Reusing on Patient
Assemble machine as described under "Preparation"
on page 29 et seq..
Carry out checks to ensure readiness for operation,
see "Device Check" on page 42.
Maintenance Intervals
Clean and disinfect equipment and/or components
before any maintenance procedures – and before
returning for repair!
O2 sensor capsule Replace sensor capsule in event
of display message:
O2 measurement inop
and if calibration is impossible.
Disposal, see page 144.
Ambient-air filter Clean or replace after 4 weeks,
Cooling-air filter see page 144.
Replace every year.
Dispose of with normal domestic
waste.
Filters in the com- To be replaced by trained service
pressed gas inlets personnel every 2 years.
Lithium battery To be replaced by trained service
for data protection personnel every 2 years.
Disposal, see page 144.
Clock module To be replaced by trained service
personnel after 6 years.
Pressure reducer Complete overhaul every 6 years
by Dräger Service.
Equipment inspection Every 6 months by trained service
and service personnel.
143
Preparing
Before Reusing on Patient
Maintenance Intervals