Dismantling
Clean and prepare the machine after each patient.
Recommendation:
Change the hose system and expiration valve every
24 hours. Keep the replacement systems ready.
Switch off the ventilator and humidifier, and remove
their power plugs.
Drain the water traps and ventilation hoses.
Drain the water container of the humidifier.
CO2 sensor (optional)
1 Remove from the cuvette. Unplug the connector from
the back of the unit.
2 Remove the cuvette of the CO2 sensor from the
Y-piece.
3 Remove the catheter cone from the cuvette.
Prepare the CO2 sensor for wipe disinfecting.
Prepare the cuvette for disinfecting and cleaning in
the autoclave.
Temperature sensor
4 Remove from the Y-piece or from the mounting of
hose set K. Do not pull the cable.
Unplug the connector from the back of the Evita 4.
Prepare the temperature sensor for wipe disinfecting.
The temperature sensor is not suitable for
autoclaving or immersion in a disinfectant bath.
1
2
3
136
Preparing
Dismantling
4
To avoid endangering hospital staff and other
patients, the ventilator must be disinfected and
cleaned whenever it has been used. Follow accepted
hospital procedures for disinfecting contaminated
parts (protective clothing, eyewear, etc.).