The following instructions include:
Equipment assembly.
Electrical and gas connections.
Setting the language for the display texts.
Automatic device check with sensor calibration.
Attaching components
Always use properly prepared parts, see Preparing,
page 135.
Fitting the expiration valve
Tilt the control unit upwards.
Push the expiration valve as far as it will go into the
mounting. Check that it is properly engaged by gently
pulling the port.
Fitting the flow sensor
1 Push socket to left as far as it will go.
2 Fit flow sensor – with the probe facing towards the
ventilator – into the mounting and push it into the
socket as far as it will go.
Then:
3 Push flow sensor to the right as far as it will go into
the rubber lip of the expiration valve.
Preparation
Attaching components
30
1
2
3