Technical Data
Environmental conditions
In operation
Temperature 10 to 40 °C
Atmospheric pressure 700 to 1060 hPa
Rel. humidity 0 to 90 %
In storage
Temperature –20 to 60 °C
Atmospheric pressure 500 to 1060 hPa
Rel. humidity 0 to 100 %
Settings
Ventilation frequency f 0 to 100/min
Inspiration time Tinsp 0.1 to 10 s
Tidal volume VT
Paediatrics 0.02 to 0.3 L, BTPS*
Accuracy ±10 % of set value, or ±10 mL,
whichever is greater.
Adults 0.1 to 2.0 L, BTPS*
Accuracy ±10 % of set value, or ±25 mL,
whichever is greater.
Inspiratory Flow
Paediatrics 6 to 30 L/min
Adults 6 to 120 L/min
Inspiratory pressure Pinsp 0 to 80 mbar
Inspiratory pressure limit Pmax 0 to 100 mbar
O2 concentration 21 to 100 Vol.%
Accuracy ±5 % of set value, or ±2 Vol.%,
whichever is greater.
Positive end-expiratory pressure PEEP or
interm. PEEP 0 to 35 mbar
Trigger sensitivity 0.3 to 15 L/min
Pressure assist PASB 0 to 80 mbar
Rise time for pressure assist 0 to 2 s
Independent lung ventilation (ILV)
Master with trigger / without trigger
Slave synchr. / asynchr. / inverse I : E
––––––––––
* BTPS = Body Temperature, Pressure, Saturated.
Measured values relating to the conditions of the patient lung:
body temperature 37 ºC, steam-saturated gas, ambient pressure.