Disinfecting/Cleaning
Use surface disinfectants. For surface compatibility, use
disinfectants based on:
aldehydes,
quaternary ammonium compounds.
To avoid the possibility of damage to material, do not use
any disinfectants based on:
alkylamine-based compounds
phenol-based compounds,
halogen-releasing compounds,
strong organic acids,
oxygen-releasing compounds.
For users in the Federal Republic of Germany, we
recommend that only disinfectants on the current DGHM
list are used (DGHM: German Society for Hygiene and
Microbiology).
The DGHM list (published by mhp-Verlag, Wiesbaden)
also classifies each disinfectant by its active agents.
For countries where the DGHM list is not available, we
recommend the types of disinfectant given above.
Disinfectants often contain – besides their main active
agents – additives that can also damage materials.
If in doubt, ask the supplier/manufacturer of the
disinfectant/ cleaning agent.
A list of substances used is given on page 158.
Do not sterilise parts in ethylene oxide!
The screen is made of Plexiglas.
Do not handle with alcohol or agents containing
alcohol. Danger of cracking.
139
Preparing
Disinfecting/Cleaning
To avoid endangering hospital staff and other
patients, the ventilator must be disinfected and
cleaned whenever it has been used. Follow accepted
hospital procedures for disinfecting contaminated
parts (protective clothing, eyewear, etc.).