95
Operation
Special Functions
Switching off the monitor functions
e.g. if a spent sensor cannot be replaced.
Immediately make sure there is an adequate external
monitor backup!
Example: Switching off Flow Monitoring.
Press »Alarm limits« key.
Display (example):
Touch »Monitoring« screen key.
Display (example):
For the example of switching off flow monitoring:
Touch the »Flow off« screen key.
The key changes colour from green to yellow.
To confirm that you wish to switch off Flow Monitoring:
Press rotary knob. Flow Monitoring is switched off,
and the corresponding measured values disappear.
The alarm function is switched off.
After replacing the sensor:
Switch the monitor function back on.