Operating data
Mains power connection 100 V to 240 V
50/60 Hz
Current
at 230 V max. 1.3 A
at 100 V max. 3.2 A
Power consumption typically approx. 125 W
Machine fuses
Range 100 V to 240 V F 5 H 250 V IEC 127-2 (2x)
Protection class
Machine Class I
CO2 sensor (sensor connected) Type BF
m
Temperature sensor AWT 01 Type BF
m
(sensor connected)
Gas supply
O2 gauge pressure 3 bar –10 % to 5.5 bar +10 %
at 60 L/min (peak flow 200 L/min)
O2 connection thread M 12 x 1, female
air gauge pressure 3 bar –10 % to 5.5 bar +10 %
at 60 L/min (peak flow 200 L/min)
air connection thread M 20 x 1.5, male
The gases must be dry and free from oil and dust.
Gas consumption of control system Medical air or O2 approx. 3.5 L/min
Output for pneumatic Medical air or O2
medicament nebuliser max. 2 bar
max. 10 L/min
Automatic gas switch-over if one gas fails (inlet pressure <1.5 bar),
the device switches to the other gas.
Sound pressure level max. 47 dB (A)
(for free-field measurement over
a reflecting surface)
Dimensions (W x H x D)
Basic machine 530 x 290 x 450 mm
Machine with trolley 580 x 1335 x 660 mm
Weight
Basic machine approx. 27 kg
Basic machine with trolley
incl. cabinet 8H approx. 69 kg
Machine outputs
Digital output Output and reception via an RS 232 C interface
COM 1 LUST protocol
Baudrate: 1200, 2400, 4800, 9600, 19200 baud
Data bits: 7
Parity: even
Stop bits: 1
MEDIBUS protocol
Baudrate: 1200, 2400, 4800, 9600, 19200 baud
Data bits: 8
Partiy: even, odd, no