BIPAPAssist
Biphasic Positive Airway Pressure Assisted
pressure-controlled, assisted ventilation
The inspiratory strokes are the same as for BIPAP, but
the changeover from Pinsp to PEEP is not synchronised
with expiration by the patient.
The patient can breathe spontaneously at PEEP level
through the entire ventilation process.
Every spontaneous breathing activity by the patient
triggers a synchronised inspiratory stroke.
A non-synchronised inspiratory stroke is started by the
device at the latest upon expiry of the time »f«.
For all patients, from those unable to breathe
spontaneously to those breathing spontaneously before
being weaned off the ventilator.
Set ventilation pattern for BIPAPAssist with the following
parameters:
Inspiratory pressure »Pinsp«
Frequency »f«
Inspiration time »Tinsp«
O2 concentration »O2«
Positive end-expiratory pressure »PEEP«
Pressure rise time » «
Flow trigger »FlowTrig«
The inspiratory pressure »Pinsp« is set as an absolute
value.
––––––––––––––––––––––––––––––––––––––––––––––––
To set:
Touch the appropriate screen knob.
Adjust to the desired value = turn rotary knob.
Confirm setting = press the rotary knob.
––––––––––––––––––––––––––––––––––––––––––––––––
Operation
Setting Ventilation Modes
62
Paw
Flow
Pinsp
PEEP
slow
rise time
fast
rise time
t
t
Tinsp
Trigger window
for synchronisation
f
1
without
spontaneous breathing
with
spontaneous breathing
081 37 207