106
Operation
Calibrating
Enter these concentration values with the screen
knobs.
Touch the relevant screen knob.
Enter the concentration = turn rotary knob.
If the calibration gas consists only of CO2 and N2, set
the O2 and N2O concentrations to 0.
Touch the »Start« screen key.
During calibration, the following message is displayed on
the screen:
CO2 calibration. Please wait
Evita 4 carries out calibration and confirms with the
message:
CO2 calibration ok
Failed calibration is indicated by the device with the
message:
CO2 calibration interrupted
or
CO2 calibration not ok
Repeat the calibration of the CO2 sensor.
Resetting CO2 calibration
If calibration with calibration gas was unsuccessful,
the factory-set calibration value can initially be used.
Press the »Calibration« key.
Touch the »CO2« screen key.
On the CO2 calibration side, touch the »Cal.« screen
key and then touch the »Alarm Reset« screen key.
After about 5 seconds, resetting is complete, and the
factory-set calibration value is active.
Recover the correct calibration as soon as possible!