Fitting CO2 cuvette and CO2 sensor
(optional)
1 Fit the cuvette to the patient connection of the
Y-piece, with the cuvette windows facing the side.
2 Push the CO2 sensor on to the cuvette, with the cable
trailing towards the unit.
Insert the probe of the CO2 sensor in the socket
»CO2 m« on the rear panel of the Evita 4.
Ventilating Infants
Up to 300 mL tidal volume VT
Patient mode »Paediatrics«
Do not use a heat and moisture exchanger at the
same time as a humidifier!
Risk of increased breathing resistance because of
condensation.
Fitting bacterial filter
Fit the bacterial filter to the inspiratory port.
1
2
Preparation
Ventilation Adults and Children
Ventilating Infants