123
Configuration
Ventilation
Setting the start-up values of the alarm limits
● Press the »Configuration« key.
● Touch the »Ventilation« screen key.
Enter the access code 3032:
● Touch the corresponding screen keys.
● Touch the »Start up settings« screen key.
● Touch the »Alarm limits« screen key.
Display (example):
Start-up values of the alarm limits:
The factory-specific start-up settings can be entered in
the table.
To change the start-up values of the alarm limits:
● Touch the screen key of the alarm limit you wish to
change.
● Change value = turn rotary knob.
● Confirm value = press rotary knob.
Alarm limit Factory settings Hospital-specific settings
Paw high [mbar] 50
......................................
MV low [L/min] (VT
●
f) –20 %
......................................
MV high [L/min] (VT
●
f) +50 %
......................................
VT high [L] VT +100 %
......................................
etCO2 low [mmHg] 30
......................................
etCO2 high [mmHg] 60
......................................
fspon [1/min] 50
......................................
TApnoea [s] 15
......................................
183 37 207