73
Operation
Independent Lung Ventilation ILV
Setting ILV/Slave
Volume-controlled ventilation with fixed, mandatory
minute volume MV, set with the tidal volume VT and
frequency f of the ILV Master device and selectable Slave
mode.
For independent lung ventilation of patients with no
spontaneous breathing.
––––––––––––––––––––––––––––––––––––––––––––––––
To set Slave mode:
● Touch the »Extra settings« screen key.
● Touch the »Slave Mode« screen key.
To select the desired slave mode (e.g. »Async.«):
● Touch the appropriate screen key and press the
rotary knob.
––––––––––––––––––––––––––––––––––––––––––––––––
Paw
Flow
T
insp Te
Pplat
Insp.
pause
Pmax
(configu-
rable)
PEEP
1
f
Insp. flow
t
t
099 37 207