EasyManua.ls Logo

Dräger Fabius plus - Restoring the factory settings

Dräger Fabius plus
232 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Configuration
142 Instructions for use Fabius plus SW 3.n
Restoring the factory settings
1 Select Restore Factory Defaults (M) and
confirm.
2 Select Yes or No (N) and confirm.
If Yes is selected, the factory settings are restored.
The factory settings replace the current default
settings.
The values for the factory settings for Fabius are
listed in the following table.
22115
Parameter Factory setting
Volume Control PMAX= 40
VT= 600
Freq= 12
TI:TE= 1:2.0
TIP:TI= 10
PEEP= 0
Pressure Control PINSP= 15
Freq= 12
TI:TE= 1:2.0
Insp Flow= 30
PEEP= 0
M
N
Standby Set-up
Pressure Support PPS= 10
Freq Min= 3
Trigger= 2
Insp Flow= 30
PEEP= 0
SIMV/PS PMAX= 40
VT= 600
Freq= 12
PPS= 10
PEEP= 0
Trigger= 2
Insp Flow= 30
TINSP= 1.7
TIP:TI= 10
Alarm limits for O
2 Upper value = 100
Lower value = 20
Alarm limits for MV Upper value = 12.0
Lower value = 3.0
Upper alarm limit and
pressure threshold for
PEAK
Upper value = 40
Lower value = 8
Minimum Alarm Vol-
ume
Volume = 5
Parameter Factory setting

Table of Contents

Other manuals for Dräger Fabius plus

Related product manuals