Instructions for use Fabius plus SW 3.n 141
Configuration
Changing the alarm limits
1 Select Alarm Limits (G) and confirm.
2 Select the alarm limits (H) to be changed and
confirm.
3 Select new value and confirm.
4 If necessary, repeat steps 2 and 3 for other
alarm limits.
5 Select the input arrow (I) and confirm.
The window is closed.
In the following table, the values for the setting
ranges and factory settings are listed for all alarm
limits of Fabius.
Changing the minimum alarm volume
1 Select Minimum Alarm Volume (K) and
confirm.
The current minimum alarm volume (L) is
displayed on the screen.
2 Set the new minimum alarm volume to a value
between 1 (minimum) and 10 (maximum) and
confirm.
The window is closed, the cursor is on the input
arrow (J).
22109
Alarm parame-
ter
Setting range Factory setting
O
2 19 to 100 100
[%] 18 to 99 20
MV 0.1 to 20.0 12.0
[L/min] 0.0 to 19.9 3.0
Pressure 10 to 70 40
[[cmH
2O ]
(hPa)]
5 to 30 8
Standby Set-up
G
H
I
22113
J
K
L
Standby Set-up