80 Instructions for use | Oxylog VE300 SW 1.n
Configuration
9.9.4 Demo mode
Demo mode disables BTPS correction so that correct values are displayed when
demonstrating a device on a test lung.
WARNING
Risk of patient injury
The demo mode changes the tidal volume.
Never activate demo mode while a patient is connected.
Here the device demo mode can be activated and deactivated.
9.9.5 Exporting data (option)
Here the following data can be exported as a log file to a USB flash drive for
analysis and documentation purposes:
System tests
Screenshots
Patient log files
The exported data can be read with common programs.
To export log files:
1. Touch the Export data button.
2. Connect a USB flash drive.
The device automatically detects the USB flash drive and transfers the saved
data to it.
A successful transfer is confirmed with the message Export successfully
completed.
3. Touch the Finish button to complete the export and remove the USB flash drive.
WARNING
Risk of patient injury and damage to the device
Do not connect a USB cable to the USB port. The voltage may cause damage to
the device or harm the patient.
Only connect USB flash drives to the USB port.
9.9.6 Changing the password
Here the password for the advanced system setup can be changed.
WARNING
Unauthorized access to the system setup
If the password is not changed, unauthorized individuals have access to the system
setup and can change important settings.
Change the preset password before the ventilator is used for the first time.