84 Instructions for use | Oxylog VE300 SW 1.n
Configuration
The preset parameter values are listed in the following table:
The preset alarm setting values are listed in the following table:
RR
[/min]
V
[mL]
I:E Ti
[s]
PEEP
[m-
bar]
Pmax
[m-
bar]
∆P-
supp
[mbar]
Trig-
ger
Slope O2/air
mix
Adult - Ventila-
tion (volume) -
VC-CMV/VC-AC
12 400 1:1.5 - 5 30 - 0 - 100 %
O
2
Adult - Ventila-
tion (volume) -
VC-SIMV
12 400 - 2.0 5 30 0 3 Me-
dium
100 %
O
2
Adult - Ventila-
tion (pressure) -
PC-BIPAP
12 - - 2.0 5 30
Pinsp
20
03Me-
dium
100 %
O
2
Adult - CPAP -
SPN-CPAP
----530- --O2/air
mix
Adult - CPAP -
SPN-CPAP/PS
----5300 3Me-
dium
O
2/air
mix
Adult - CPR -
VC-CMV
10 400 1:1.5 - 0 60 - - - 100 %
O
2
Child - Ventila-
tion (volume) -
VC-CMV/VC-AC
20 200 1:1.5 - 5 25 - 0 - 100 %
O
2
Child - Ventila-
tion (volume) -
VC-SIMV
20 200 - 1.2 5 25 0 3 Me-
dium
100 %
O
2
Child - Ventila-
tion (pressure) -
PC-BIPAP
20 - - 1.2 5 25
Pinsp
15
03Me-
dium
100 %
O
2
Child - CPAP -
SPN-CPAP
----525- --O2/air
mix
Child - CPAP -
SPN-CPAP/PS
----5250 3Me-
dium
O
2/air
mix
Child - CPR -
VC-CMV
10 200 1:1.5 - 0 40 - - - 100 %
O
2
Patient category Alarm limit Value
Adult MVe low - MVe high 3 - 10 L/min
RR high 30 /min
etCO
2 low - etCO2 high 30 - 50 mmHg /
4.0 - 6.7 kPa /
4.0 - 6.7 Vol%
etCO
2 low - etCO2 high
(only CPR)
0 - 99 mmHg /
0 - 13.2 kPa /
0 - 13.2 Vol%