STP-iXR1606 Turbomolecular Pump
MT-79E-001-D
Page 142
7
SAFETY FUNCTIONS
Referred
Section
Table 43 - Error List (continued)
Countermeasures
Set the ambient
temperature to 40 °C or
less.
Set the cooling water flow
rate at 3 L/min and cooling
water temperature
Set the temperature of the
baking heater to 120 °C or
less.
Always cool the STP pump.
Check the piping system
for leakage.
Repetitious start/stop
operations may cause the
STP pump to overheat.
heck the exhaust sequence.
switch
the pump, and then turn "ON"
switch again to recover.
Probable causes
High ambient temperature.
Insufficient cooling of the
control unit.
Overheating during baking.
Insufficient cooling.
Leakage of the vacuum
piping system.
Repetitive start/stop
operations.
Atmospheric air flows the STP
.