1.9 Vapor Recovery Application
Converter effi ciency
The measurement accuracy is highly depen-
dent on the converter effi ciency: If this is too
low, the converter material needs replace-
ment.
To measure the converter effi ciency, one
has to compare the measurement values of
CH1 with and without having the gas fl owing
through the converter. This requires activating
the valve V1.
4 Maintenance section of this manual for
instructions about when and how to replace
the converter material.