Emerson Process Management GmbH & Co. OHG4-2
X-STREAM X2FD
Instruction Manual
HASXMDE-IM-EX
01/2015
4 Service and Maintenance
FLAMMABLE GASES - EXPLOSION HAZARD
Leaks may cause explosion when measuring ammable gases!
When measuring ammable gases it is recommended to perform a leak test
on all gas paths, connections and components before startup or applying
power. Leak tests should be carried out on a 2 month‘s regular basis and
after repair/maintenance.
See the analyzer instruction manual for instructions on how to carry out
leak tests.
HAZARDOUS GASES
When measuring ammable and / or toxic gases it is recommended to
purge the system with air or an inert gas, e.g. nitrogen, prior to opening
the gas paths.
Violation may cause an explosion and/or personal injury or death!
EXPLOSION HAZARD
Modi cations affecting the integrity of type of protection (e.g. af xing
addtional threads, replacing the ame arrestors by other model) are NOT
PERMITTED!
Violation may cause an explosion and/or personal injury or death!