4 Service and Maintenance
FLAMMABLE GASES - EXPLOSION HAZARD
Leaks may cause explosion when measuring fl ammable gases!
When measuring fl ammable gases it is recommended to perform a leak test
on all gas paths, connections and components before startup or applying
power. Leak tests should be carried out on a 2 month‘s regular basis and
after repair/maintenance.
See the analyzer instruction manual for instructions on how to carry out
leak tests.
HAZARDOUS GASES
When measuring fl ammable and / or toxic gases it is recommended to
purge the system with air or an inert gas, e.g. nitrogen, prior to opening
the gas paths.
Violation may cause an explosion and/or personal injury or death!
EXPLOSION HAZARD
Modifi cations affecting the integrity of type of protection (e.g. affi xing
addtional threads, replacing the fl ame arrestors by other model) are NOT
PERMITTED!
Violation may cause an explosion and/or personal injury or death!