Emerson Process Management GmbH & Co. OHG5-20
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Signal: Comp1
OutRange: 0-20 mA
Low Scale: 0.00
Max Scale: 100.00
AutoScale: Yes
FailMode: High +10%
0/4 mA: 0.00
20 mA: 100.00
Hold:
No
5.7 Checking the Settings
Next, select the output range:
0-20 mA (dead zero) generates a 20 mA
signal, if a concentration is measured at
the upper limit of the signal range. A 0 mA
signal is generated if the sample gas con-
centration equals the value specied with
"LowScale" .
4-20 mA (life zero): A 4 mA signal is gene-
rated if the concentration equals the value
specied with "LowScale", thus enabling to
detect e.g. a broken cable.
"FailMode" selects the output´s behaviour under
failure conditions, considering or not, the NA-
MUR recommendation NE 43. NE 43 denes
output signals enabling to detect different types
of failures/status ( table 5-1): The related
information is transmitted as a current signal,
but outside the (0)4-20 mA measurement signal
range.
Available options:
Track: NE 43 not considered.
HIGH +10%: NE 43 failure signal level:
"above".
LOW -10%: NE 43 failure signal level:
"below".
Note!
The related outputs signals can be netuned
on a second menu page, if "FailMode" is set
to other than Track ( next section).
Enter a concentration, to equal the low output
limit (0 or 4 mA)
Enter a concentration, to equal the high output
limit (20 mA)
Enable (Yes) or disable (No) output autosca-
ling.
Note!
Factory setting is "Outrange:" 4-20 mA and
"FailMode:" LOW - 10%, unless ordered
otherwise.