Emerson Process Management GmbH & Co. OHG6-78
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Serial
Protocol: Modbus-RTU
32Bit Mode: 16BitLow
Mbus ID: 1
Baud Rate: 19200
Parity: Even
6.2.3.5.1 Setup Communication Serial
Setup..
Communication..
Serial..
6.2.3 Setup Menu
Select the Modbus protocol.
Available options:
MODB-RTU, None
Modbus mode of operation.
Available options:
32Bit (=Daniel mode),
16BitLow (=Modicon mode, LOW word rst)
16BitHigh (=Modicon mode, HIGH word rst)
Enter instrument ID for network.
Accepted values: 1 .. 254
Select baud rate for the serial interface.
Available options:
2400, 4800, 9600, 19200
Set whether a parity bit is used.
Available options:
None, Even, Odd