Maintenance & Procedures
From the table below, in the rst column select
your preferred the calibration procedure, and
notice the information in the columns aside.
7.4 Calibration Procedures
Type of Procedure
Menu Page
(CONTROL - )
Valves
Simulta-
neously
Calibrated
Channels
More Information
Manual calibration
ZERO CALIBRATION...
optional single channel page 7-18
SPAN CALIBRATION...
Advanced calibration
ADV.CALIBRATION - ZEROALL!
required all channels page 7-21ADV.CALIBRATION - SPANALL!
ADV.CALIBRATION - ZSCALALL
!
Remote calibration n.a. (via Modbus or Dig IN) recommended all channels page 7-35
Unattended calibration n.a. (via interval time) required all channels page 7-39
Proper conguration and performing of calibrations is essential to keep the
functionality of your analyzer. Therefore, to avoid miscalibrations, several menus
can be locked by access codes.
Descriptions in subsequent sections do not care about locking of menus.
Information about locking menus are provided in chapter 6.