Emerson Process Management GmbH & Co. OHG7-46
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Don‘t forget now to set „Activa-
ted“ to the previous value after
nishing the cross interference
compensation setup!
Ch1
IF Source1
Source: None
Value 0
Status Good
Interference Factor
: 1
Adjust IF Factor!
Linearization Curve..
6a. If the signal of the selected source is
linear:
To automatically calculate the proper
interference compensation factor, enter
Adjust IF Factor!“ and press enter.
The analyzer now starts calculating and,
when nished, shows the result in line
Interference Factor:
Note!
If it turns out during measurement, that
the calculated factor is not satisfactory,
you may change the „Interference Factor
manually. Simply enter the line, press
enter and change the value.
7.5 Cross Interference Compensation
7. If need be,
repeat from step 4 for more sources of
interference for the currently selected
channel,
or
start from step 1 to congure interference
compensation for another channel.