Emerson Process Management GmbH & Co. OHG 8-7
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
8
Troubleshooting
Message
Status level
Explanation Recommended Actions
USB free space warning
Maintenance
The attached USB storage
devices´s free memory ex-
ceeded the setup limit (
6-103 )
Replace USB device by ano-
ther one with sufcient free
memory
On the USB device, delete
les not required. To do so,
attach it to a computer.
Format USB device (
6-104 )
USB stick full
Maintenance
The attached USB storage
device has not sufcient free
memory to store data