EasyManua.ls Logo

Emerson X-STREAM XE - Page 410

Emerson X-STREAM XE
536 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Emerson Process Management GmbH & Co. OHG8-40
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
8.4 Troubleshooting on Components
R60
Test connector P4
P4.15
P4.16 +
_
Step 3:
Locate test connector P4 to measure
the bridge voltage:
P4.16 Bridge voltage (+)
P4.15 Bridge voltage (-); GND
CAUTION!
Do not short-circuit pins!
Alternatively the GND signal (-) is ac-
cessible on the main board BKS, too:
Locate X11 (
g. 8-3, page 8-16) .
The bridge voltage depends on range
and sample gas and should be bet-
ween 3V and 5V.
Only if the WAP 100 board has been
replaced, it is necessary to adjust the
voltage with potentio-meter R60.

Table of Contents