9.3 List of Parameters and Registers - Sorted by Tag Name
Tag Name Address
Data
Type
Client
Access
Description
Modicon Daniel
Control.ApplyGas.Span2Valve4 2074 2074 Boolean R/W
0=open sample valve; 1=open span2
valve comp4
Control.ApplyGas.Span2Valve5 2078 2078 Boolean R/W
0=open sample valve; 1=open span2
valve comp5
Control.ApplyGas.Span3Valve1 2063 2063 Boolean R/W
0=open sample valve; 1=open span3
valve comp1
Control.ApplyGas.Span3Valve2 2067 2067 Boolean R/W
0=open sample valve; 1=open span3
valve comp2
Control.ApplyGas.Span3Valve3 2071 2071 Boolean R/W
0=open sample valve; 1=open span3
valve comp3
Control.ApplyGas.Span3Valve4 2075 2075 Boolean R/W
0=open sample valve; 1=open span3
valve comp4
Control.ApplyGas.Span3Valve5 2079 2079 Boolean R/W
0=open sample valve; 1=open span3
valve comp5
Control.ApplyGas.Span4Valve1 2064 2064 Boolean R/W
0=open sample valve; 1=open span4
valve comp1
Control.ApplyGas.Span4Valve2 2068 2068 Boolean R/W
0=open sample valve; 1=open span4
valve comp2
Control.ApplyGas.Span4Valve3 2072 2072 Boolean R/W
0=open sample valve; 1=open span4
valve comp3
Control.ApplyGas.Span4Valve4 2076 2076 Boolean R/W
0=open sample valve; 1=open span4
valve comp4
Control.ApplyGas.Span4Valve5 2080 2080 Boolean R/W
0=open sample valve; 1=open span4
valve comp5
Control.ApplyGas.ZeroValve1 2056 2056 Boolean R/W
0=open sample valve; 1=open zero
valve comp1
Control.ApplyGas.ZeroValve2 2057 2057 Boolean R/W
0=open sample valve; 1=open zero
valve comp2
Control.ApplyGas.ZeroValve3 2058 2058 Boolean R/W
0=open sample valve; 1=open zero
valve comp3
Control.ApplyGas.ZeroValve4 2059 2059 Boolean R/W
0=open sample valve; 1=open zero
valve comp4
Control.ApplyGas.ZeroValve5 2060 2060 Boolean R/W
0=open sample valve; 1=open zero
valve comp4
Control.Calibration.Blowback_1 2016 2016 Boolean R/W Blowback procedure comp1 (1=start)
Control.Calibration.Blowback_2 2017 2017 Boolean R/W Blowback procedure comp2 (1=start)
Control.Calibration.Blowback_3 2018 2018 Boolean R/W Blowback procedure comp3 (1=start)
Control.Calibration.Blowback_4 2019 2019 Boolean R/W Blowback procedure comp4 (1=start)
Control.Calibration.Blowback_5 2020 2020 Boolean R/W Blowback procedure comp5 (1=start)
Control.Calibration.Blowback_All 2025 2025 Boolean R/W Blowback procedure all (1=start)
Control.Calibration.Calib_Cancel 2026 2026 Boolean R/W Cancel any calibration (1=cancel)
Control.Calibration.ProgSequence 2024 2024 Boolean R/W Zero+span calibration all (1=start)
Control.Calibration.Span_1 2006 2006 Boolean R/W Span calibration comp1 (1=start)
Control.Calibration.Span_2 2007 2007 Boolean R/W Span calibration comp2 (1=start)
Control.Calibration.Span_3 2008 2008 Boolean R/W Span calibration comp3 (1=start)
Control.Calibration.Span_4 2009 2009 Boolean R/W Span calibration comp4 (1=start)