PROline Promag 53 9 Troubleshooting
Endress+Hauser 101
S
$
SENSOR SW DAT
# 032
Sensor:
Error accessing the calibration
values stored in the S-DAT™.
1. Check whether the S-DAT™ is
correctly plugged into the
amplifier board
Page 113,
115
2. Replace the S-DAT™ if it is
defective. Spare parts
Page 111
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
Spare part set number
Hardware revision code
3. Replace measuring electro-
nics boards if necessary.
Spare parts
Page 111
S
$
TRANSM. HW-DAT
# 041
Transm itter DAT:
1. T-DAT™ is defective.
2. T-DAT™ is not plugged into
the amplifier board or is
missing.
1. Replace the T-DAT™.
Spare parts
Page 111
Check the spare part set
number to ensure that the new,
replacement DAT is compati-
ble with the measuring elec-
tronics.
2. Plug the T-DAT™ into the
amplifier board
Page 113,
115
S
$
TRANSM. SW-DAT
# 042
Transm itter DAT:
Error accessing the calibration
values stored in the T-DAT™.
1. Check whether the T-DAT™ is
correctly plugged into the
amplifier board
Page 113,
115
2. Replace the T-DAT™ if it is
defective.
Spare parts
Page 111
Before replacing the DAT,
check that the new, replace-
ment DAT is compatible with
the measuring electronics.
Check the:
Spare part set number
Hardware revision code
3. Replace measuring electro-
nics boards if necessary.
Spare parts
Page 111
S
$
A / C COMPATIB.
# 051
The I/O board and the amplifier
board are not compatible.
Use only compatible modules and
boards. Check the compatibility of
the modules used.
Check the:
Spare part set number
Hardware revision code
S
$
HW F-CHIP
# 061
Transmitter F-Chip™:
1. F-Chip™ is defective.
2. F-Chip™ is not plugged into
the I/O board or is missing.
1. Replace the F-Chip™.
Accessories
Page 97
2. Plug the F-Chip™ into the I/O
board
Page 113, 115
Nr. # 1xx
Software error
S
$
CHECKSUM TOTAL
# 111
Totalizer checksum error 1. Restart the measuring device
2. Replace the amplifier board if
necessary.
Spare parts
Page 111
Type Error message / No. Cause Remedy / spare part