10 Technical data PROline Promag 53
148 Endress+Hauser
Promag H / 1/12" to 1" / DN 2 - 25 (remote version)
Fig. 75: Dimensions Promag H / 1/12" to 1" / DN 2 - 25 (remote version)
Dimensions of wall-mount housing
→ see Page 136
Wall mounting set for Promag H / 1/12" to 1" / DN 2 - 25
Fig. 76: Wall mounting set for Promag H / 1/12" to 1" / DN 2 - 25
A = 4.92" (125 mm), B = 3.46" (88 mm), C = 4.72" (120 mm), D = Ø 0.28" (7 mm)
DN PN * DI L A B C K M
DIN
[mm]
ANSI
[inch]
DIN
psig (bar)
inch
(mm)
inch
(mm)
inch
(mm)
inch
(mm)
inch
(mm)
inch
(mm)
[mm]
2 1/12 230/580 (16/40) 0.09 (2.25) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 1.69 (43) M 6x4
4 5/32 230/580 (16/40) 0.18 (4.5) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 1.69 (43) M 6x4
8 5/16 230/580 (16/40) 0.35 (9.0) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 1.69 (43) M 6x4
15 1/2 230/580 (16/40) 0.63 (16.0) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 1.69 (43) M 6x4
– 1 230/580 (16/40) 0.89 (22.6) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 2.09 (53) M 6x4
26 − 230/580 (16/40) 1.02 (26.0) 3.39 (86) 8.39 (213) 5.83 (148) 2.56 (65) 2.09 (53) M 6x4
Overall fitting length will depend on the process connections → Page 151
* The permissible nominal pressure depends on the process connection and seal:
– 580 psig/40 bar: flange, welded nipples (with O-ring seal)
– 230 psig/16 bar: all other process connections
F06-xxHxxxxx-06-05-xx-xx-001ae
Dimensions in inches (mm)
A
B
C
D
F06-xxHxxxxx-06-07-06-xx-000