PROline Promag 53 3 Installation
Endress+Hauser 33
Assembly with ground disks (1/2" to 12" / DN 15 to 300)
Depending on the application, e.g. with lined or ungrounded pipes (see Page 56), it
may be necessary to mount ground disks between the sensor and the pipe flange for
the potential equalization. Ground disks can be ordered separately as an accessory
from E+H (see Page 97).
"
Caution:
• In this case, when using ground disks (including seals) the total fitting length
increases! Dimensions → Page 146.
• PTFE and PFA lining → only install additional seals between the ground disk and
pipe flange.
1. Place the ground disk and the additional seal between the instrument and the pipe
flange (Fig. 18).
2. Insert the bolts through the flange holes. Tighten the nuts so that they are still loose.
3. Now rotate the ground disk as shown in Fig. 18 until the handle strikes the bolts.
This will center the ground disk automatically.
4. Now tighten the bolts to the required torque (see Page 35)
5. Connect the ground disk to ground → Page 58.
Fig. 18: Assembly with ground disks (Promag P, 1/2" to 12" / DN 15 to 300)
F06-5xFxxxxx-17-05-xx-xx-001