Proline Promass 83 PROFIBUS DP/PA Trouble-shooting
Endress+Hauser 125
206 S SAVE T-DAT
! # 206
0x0F BAD Device Failure Constant Restore from T-
DAT failed
Cause:
System error. Data backup to T-DAT failed, or error
when accessing the calibration values stored in the
T-DAT.
Remedy:
1. Verify that the T-DAT is correctly plugged into
the amplifier board, Page 139 ff. Replace T-
DAT if defective. Spare parts Page 137 ff.
Check the spare part set number to ensure that
the new, replacement DAT is compatible with
the measuring electronics.
2. Replace measuring electronics boards if
necessary.
Spare parts Page 137 ff.
251 S: COMMUN. SENS.
$: # 251
0x0F BAD Device Failure Constant Communication
Sensor Failure
Cause:
Internal communication fault on the amplifier
board.
Remedy:
Remove the amplifier board. Spare parts
Page 137 ff.
261 S COMMUNICAT. I/O
$ # 261
0x18
0x19
0x1A
BAD No Commu-
nication
O.K.
Low
High
Communication
failure
Cause:
System error. Communication error. No data
reception between amplifier and I/O board or faulty
internal data transfer.
Remedy:
Check whether the electronics boards are correctly
inserted in their holders Page 139 ff.
339
...
342
S: STACK CUR OUT n
$: # 339...342
0x0F BAD Device Failure Constant Temp. Buf. not
cleared
Cause:
The temporarily buffered flow portions (measuring
mode for pulsating flow) could not be cleared or
output within 60 seconds.
Remedy:
1. Change the upper or lower limit setting, as
applicable.
2. Increase or reduce flow, as applicable.
Recommendations in the event of fault category =
FAULT MESSAGE ($):
Configure the fault response of the output to
“ACTUAL VALUE”, so that the temporary buffer
can be cleared. Page 137
Clear the temporary buffer by the measures
described under Item 1.
343
...
346
S: STACK FREQ. OUT n
$: # 343...346
0x0F BAD Device Failure Constant Temp. Buf. not
cleared
No. Device status message
(local display)
PROFIBUS measured value status
Adv. diagnosis
message in
PROFIBUS
master
Cause/remedy
Quality code (HEX)
Quality status
Quality substatus