Trouble-shooting Proline Promass 83 PROFIBUS DP/PA
128 Endress+Hauser
382 S FLUIDTEMP.MAX.
$ # 382
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
Fluid
Temperature
Max.
Cause:
The temperature sensor on the measuring tube is
likely defective.
Remedy:
Check the following electrical connections before
you contact your E+H service organisation:
Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
Page 139 ff.
Remote version: Check sensor and transmitter
terminal connections No. 9 and 10
Page 30 ff.
383 S CARR.TEMP.MIN
$ # 383
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
Carrier
Temperature
Min.
Cause:
The temperature sensor on the carrier tube is likely
defective.
Remedy:
Check the following electrical connections before
you contact your E+H service organisation:
Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
Page 139 ff.
Remote version: Check sensor and transmitter
terminal connections No. 11 and 12
Page 30 ff.
384 S CARR.TEMP.MAX
$ # 384
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
Carrier
Temperature
Max.
Cause:
The temperature sensor on the carrier tube is likely
defective.
Remedy:
Check the following electrical connections before
you contact your E+H service organisation:
Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
Page 139 ff.
Remote version: Check sensor and transmitter
terminal connections No. 11 and 12
Page 30 ff.
385 S INL.SENS.DEF.
$ # 385
0x10
0x11
0x12
BAD Sensor Failure O.K.
Low
High
Inlet Sensor
Defect
Cause:
One of the measuring tube exciter coils (inlet or
outlet) is likely defective.
Remedy:
Check the following electrical connections before
you contact your E+H service organisation:
Verify that the sensor signal cable connector is
correctly plugged into the amplifier board
Page 139 ff.
Remote version: Check sensor and transmitter
terminal connections No.4,5, 6,
7 Page 139 ff..