Commissioning Proline Promass 83 PROFIBUS DP/PA
92 Endress+Hauser
!
Note!
The measured variables for the software options (concentration measurement, viscosity, batching,
adv. diagnosis) are only available if the appropriate additional software is installed in the device. If
the measured variable is selected even if the additional software is not installed, the value “0” is
transmitted to the automation device (PROFIBUS master) for the measured variable.
Example:
You want to cyclically transmit the mass flow to the automation device by means of the Analog
Input function block 1 (module AI) and the temperature by means of the Analog Input function
block 2.
Configuration via the local display:
1. BASIC FUNCTION PROFIBUS DP FUNCTION BLOCKS BLOCK SELECTION :
Select ANALOG INPUT 1, then select CHANNEL =MASS FLOW
2. BASIC FUNCTION PROFIBUS DP FUNCTION BLOCKS BLOCK SELECTION :
Select ANALOG INPUT 2, the select CHANNEL = TEMPERATURE
Configuration via the operating program:
1. Open Analog Input function block 1 and enter the ID for mass flow in the “CHANNEL”
parameter 277.
2. Open Analog Input function block 2 and enter the ID for temperature in the “CHANNEL”
parameter 285.
°Baume
1175
°API
°Plato
°Balling
°Brix
The following measured variables are available when the additional “viscosity” software is installed (order option)
Dyn. viscosity 1177
Kin. viscosity 1178
T. comp. dyn. viscosity 1179
T. comp. kin. viscosity 1180
The following measured variables are available when the additional “batching” software is installed (order option)
Batch upwards 1346
Batch downwards 1347
Batch counter 1348
Total batching quantity 1349
Batching time 1354
The following measured variables are available when the additional “advanced diagnostics” software is installed (order
option)
Mass flow deviation 1603
Density deviation 1620
Reference density deviation 1637
Temperature deviation 1654
Tube damping deviation 1671
Actual deviation electr. dyn. sensors 1688
Measured variable ID for “CHANNEL” parameter