Commissioning Proline Promass 83 PROFIBUS DP/PA
96 Endress+Hauser
CONTROL_BLOCK module - cyclic control of device functions
By means of the CONTROL_BLOCK module, the measuring device is able to process device-specific
control variables in cyclic data exchange (e.g. switching on positive zero return).
Data structure of the CONTROL_BLOCK module:
The table describes the possible control variables which are supported by the CONTROL_BLOCK
module:
Byte 1
Control
Module Control variables for totalizer 1...3
CONTROL_BLOCK
0 → 2: Positive zero return ON
0 → 3: Positive zero return OFF
0 → 4: Zero point adjustment
0 → 8: Unidirectional (measuring mode)
0 → 9: Bidirectional (measuring mode)
0 → 24: Run SET UNIT TO BUS function
The following control variables are only executed if the measuring device has a relay
output.
0 → 50: Relay output 1 OFF
0 → 51: Relay output 1 ON
0 → 55: Relay output 2 OFF
0 → 56: Relay output 2 ON
The following control variables are only executed if the additional “batching” software
is installed in the measuring device (order option).
0 → 30: Select batching specification 1
0 → 31: Select batching specification 2
0 → 32: Select batching specification 3
0 → 33: Select batching specification 4
0 → 34: Select batching specification 5
0 → 35: Select batching specification 6
0 → 40: Stop batching
0 → 41: Start batching
0 → 42: Interrupt batching
0 → 43: Continue batching
0 → 44: Reset the fault message (affects the following batching fault messages: # 471,
472, 473, 474)
0 → 46: Reset batching quantity counter and total batching quantity to 0
The following control variables are only executed if the additional “concentration
measurement” software is installed in the measuring device (order option).
0 → 60: Select concentration specification 1
0 → 61: Select concentration specification 2
0 → 62: Select concentration specification 3
0 → 63: Select concentration specification 4
The following control variables are only executed if the additional “advanced
diagnostics” software is installed in the measuring device (order option).
0 → 25: Warning mode “Advanced Diagnostics” OFF
0 → 26: Warning mode “Advanced Diagnosis” ON
0 → 70: Start determining the user reference status
0 → 74: Acquisition mode OFF
0 → 75: Acquisition mode, periodical
0 → 76: Acquisition mode, manual
0 → 78: Start manual determination of the advanced diagnostics parameters