Installation
Hardware Installation and Maintenance Manual 3
13
5. Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal blocks (X1).
6. Attach the neutral and hot wires to the power terminal blocks by
connecting the neutral wire to terminal X1-2, the hot wire to
terminal X1-1, as shown in Figure 3–6 on page 3–14.
7. Connect the ground wire to the ground bus bar marked .
8. Verify that each connection is secure.
9. Close the analyzer enclosure cover according to the procedure under
“To close the analyzer enclosure cover” on page 3-3.
Hazardous voltage and risk of electric shock. Failure to
properly ground the analyzer may create a high-voltage shock
hazard.
Ensure that the conductors of different cross-sectional areas are
not connected into one terminal unless they are first secured with
a single compression-type ferrule. Also, to avoid risk of a short
circuit between adjacent connectors in terminal blocks, ensure
each wire uses a single compression-type ferrule.