SS2100i-1 Gas Analyzer
B18 4900002224 rev. E 12-18-20
18. Place the pressure sensor harness cap on the top of the pressure
sensor over the pins, applying light pressure until the cap snaps in
place. Refer to Figure B–12 below.
19. Secure the black harness cap on the pressure sensor by tightening
the screw on the top of the cap using a 1/8” x 2-1/2” flat-head
screwdriver.
20. Replace the upper level panel making sure to keep wires in channels
so that they are not pinched between the panel and the enclosure
sides. Refer to “Reconnect cables on the electronics assembly
panel.” on page B-8.
21. Close the analyzer enclosure cover according to the procedure under
“To close the analyzer enclosure cover” on page 3-3.
22. Conduct a leak test to determine that the new pressure sensor is not
leaking.
Do not force connection or pins may be damaged.
Make sure the black connector at the top of the pressure sensor
is facing parallel to the junction tubing to facilitate connection.
Refer to Figure B–12 above.
Do not allow cell to exceed 10 PSIG or damage could occur.
Figure B–12 New pressure sensor installed

Table of Contents

Other manuals for Endress+Hauser SpectraSensors SS2100i-1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors SS2100i-1 and is the answer not in the manual?

Endress+Hauser SpectraSensors SS2100i-1 Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors SS2100i-1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals