SS2100i-1 Gas Analyzer
A–2 4900002224 rev. E 12-18-20
Exd Accessory Conditions of Use
All accessories listed in Table A–2 below shall comply with the latest IEC/EN
60079-0 and IEC/EN 60079-1 in addition to the following conditions:
Table A-1 SS2100i-1 analyzer specifications (Continued)
Physical Specifications
Electronics Enclosure IP66 copper-free aluminum with marine
environment RAL 7001 gray polyurethane enamel
finish; approximately 300 m final thickness
Analyzer Dimensions
1
1. Consult system drawings for analyzers with sample conditioning systems.
670 mm H 580 mm W 377 mm D
(26.3” H 23” W 14.8” D)
Analyzer Weight
1
Approximately 86 kg (190 lbs)
Sample Cell Construction 316L series polished stainless steel - Standard
Number of Sample Cells 1 per analyzer
Analyzer Certifications
Certification
2
2. Including an “X” after the certificate number indicates that the product is subject to the
Special Conditions of Use specified in the schedule to this certificate. Refer to the
Special Conditions of Use under the relevant certification prior to installation or use.
CE, ATEX and IECEx: II 2 G, Ex db IIB+H2
T4 GB; LCIE 11 ATEX 3071X, IECEx LCI 11.0057X
Analyzer Assembly II 2 G, IIB+H2 T3 Gb
For a complete listing of new or updated certificates, please visit
the product page at www.spectrasensors.com.
Table A–2 Exd accessory conditions of use
Accessory
Type
Rating Notes
Stopper Plug Exd, Zone 1 1. Stopping plugs shall be assembled in such a way
that their protrusion from an associated
enclosure is not increased.
2. Installer must ensure the stopping plug ingress
protection matches the ingress protection rating
of the associated enclosure, IP66.