10 • Maintenance and servicing
108 / 144
Art.-Nr.: 80116-271
04.06
Fig. 10-7
• The test specimen is connected to the power supply via the power
cord.
• The test specimen is connected to the testbox.
• On the test specimen the test program "TP generator" is selected.
In the test program the output receptacle at which measurement
takes place is selected (in this case "Output receptacle:
2
MONOPOLAR").
Test procedure
1. On the Testbox press button S1.
2. In the test program read off the measured value for "Spark". The
tolerance range is 245 to 285 ERBE.