8 • Measurement and adjustment
Art.-Nr.: 80116-271
04.06
CHAPTER 8
Measurement and adjustment
Measurement of the HF power output
Temperature conditions
Test equipment
IMPORTANT! Measurement/adjustment should take place when the unit has
warmed up (standby temperature). For this purpose, switch on the
unit and wait until the temperature attains a constant value. This
value is between +45
°C and +55 °C and is attained after approx.
half an hour. (This value is displayed in the test program "TP gen
-
erator", for example.)
IMPORTANT! The following list contains the testing and measuring equipment
recommended by ERBE for servicing. Where ERBE article num
-
bers are specified, only original ERBE testing and measuring
equipment should be used.
ERBE
Art. No.
Description
– HF power meter
(recommended: Metron QA-ES)
20190-045 Elektrode handle ICC/ACC
20189-101 Dual-pedal footswitch with ReMode
20192-127
20192-110
Patient cable AE
or
Patient cable AE, international
20194-070
20194-075
Patient cable NE
or
Patient cable NE, international
20196-045
20196-053
Bipolar cable
or
Bipolar cable, international