7 • Test programs
Art.-Nr.: 80116-271
04.06
Test program "Measured values" (only with V 1.3.x
and later)
Fig. 7-10
Low voltage It is not possible to measure all the operating voltages simultaneous-
ly. All the blue measured values are being measured at the present
moment. All the gray values are those last measured, and they are
not updated.
To change the measured value range, you must press the selection
button next to the measured value you require.
Other measured values
U DC: output voltage of high-voltage power supply unit
I DC: output current of high-voltage power supply unit
UpHF: peak value (pos.) of HF voltage
IpHF: peak value of HF current
Phase: phase angle (0...90°)
Spark: value measured by spark sensor
(no spark = 44 ERBE)
UHF red.: value measured by redundant voltage sensor
Tem p. : Inside temperature of unit
(measured on “CPU + Sensors” circuit board)