9 • Troubleshooting
Art.-Nr.: 80116-271
04.06
B-Errors
Recognizing module: B = Control panel
Recog-
nizing
module
Error
code
Additional informa-
tion
Description Remedy
B 1 Software watchdog:
Occurs if the program
crashes due to a software
error or malfunction (e.g.
EMC).
One-off event -> EMC prob-
lem => Check environment.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Ser
-
vice.
B 2 Bit combination con-
sisting of the nonap-
plication module
codes (hexadecimal):
HF module
0X00000001
APC module
0X00000002
IES module
0X00000004
master remote control
0X00000010
slave remote control
0X00000020
1 pedal footswitch
0X00000100
2 pedal footswitch
0X00000200
multifunctional foot-
switch 0X00000400
VIO receptacle 1
0X00010000
VIO receptacle 2
0X00020000
VIO receptacle 3
0X00040000
NE receptacle
0X00080000
APC receptacle 1
0X00100000
APC receptacle 2
0X00200000
AutoStart monitor
0X10000000
Timeout monitoring:
Occurs if a module involved
in activation (e.g. HF mod
-
ule, footswitch, APC mod-
ule) fails to transmit a valid
status message via CAN for
longer than 110
ms.
Check module given in
additional information.
Communication problem
due to EMC interference
possible => Check environ
-
ment.