8 • Measurement and adjustment
Art.-Nr.: 80116-271
04.06
Adjustment
Personnel requirements
Temperature conditions
Test sequence
ATTENTION! Adjustments, technical tests, modifications, maintenance and
repair work may only be performed by ERBE or persons trained by
ERBE. If the work is not performed by trained persons, ERBE
accepts no liability and warranty rights become void.
IMPORTANT! Measurement/adjustment should take place when the unit has
warmed up (standby temperature). For this purpose, switch on the
unit and wait until the temperature attains a constant value. This
value is between +45
°C and +55 °C and is attained after approx.
half an hour. (This value is displayed in the test program "TP gen
-
erator", for example.)
IMPORTANT! Please ensure you follow the sequence specified for the individual
tests in this chapter.