7 • Test programs
Art.-Nr.: 80116-271
04.06
Exit Test programs
1. Keep pressing the Up button (9) until "Restart" appears in the
display.
2. Confirm the setting with the Enter button (12). The unit exits the
test program mode.
Parameter inputs in Test programs
In some test programs it is possible to select or change values (e.g.
for power supply unit voltage).
Example 1. Start a test program.
2. Press the selection button (1..8) next to the value to be selected
or changed. The value is shown in red and can now be changed.
3. Set the required value using Up/Down buttons (9/10).
4. Confirm value set. To do so press the selection button (1..8) next
to the value set again. The value is now shown in black again.
5. Use the Up/Down buttons (9/10) to switch to another test pro-
gram.