10 • Maintenance and servicing
Art.-Nr.: 80116-271
04.06
User manual and visual inspections
Test specimen and accessories (where enclosed) undamaged
externally.
User manual present.
All labels on the test specimen (conformity declaration mark, rat-
ing plate, and all wording) present and readily legible.
Tests to be conducted in accordance with the
national specifications and regulations
Grounded conductor test Ground terminal to chassis.
Ground terminal to potential equalization pin.
Leakage current
measurement
Ground leakage current, normal condition (N.C.).
Ground leakage current, single-fault condition (S.F.C.).
IMPORTANT! For the following tests close the output relays of the unit
with the “TP relay” test program.
Patient leakage current, normal condition (N.C.).
Patient leakage current, single-fault condition (S.F.C.).
DC resistance
Testing and measuring equipment
ERBE
Art. No.
Description
20192-127
20192-110
Patient cable AE
or
Patient cable AE, international
20190-045 Electrode handle ICC/ACC
20194-070
20194-075
Patient cable NE
or
Patient cable NE, international
20100-033 Adapter cable NE
Safety tester
(with insulation testing >500 V DC)

Table of Contents