6 • SET-UP
36 / 144
Art.-Nr.: 80116-271
04.06
Call up SET-UP
Fig. 6-1
SET-UP level 1 1. Call up "Guide" window.
2. Select menu item "Other functions".
3. Select menu item "Setup". The unit switches to SET-UP level 1.
See above table for settings that can be changed here.
SET-UP level 2 1. Call up SET-UP level 1 as described above.
2. Use the Down button (10) to scroll to the setting "Service pro-
gram".
3. Select setting "Service program".
4. Enter VIOD as the password:
Use the Up/Down buttons (9/10) to select the letters, confirming
each of the four letters with the adjacent selection button and
then jumping forward to enter the next letter. Repeat this proce
-
dure until all four letters have been entered.
5. Confirm the complete password using the Enter button (12). The
unit switches to SET-UP
level 2. See above table for settings that
can be changed here.
IMPORTANT! As from Version 1.3.x there are various methods of scrolling for-
wards within a menu:
(a) with the Down button
or
(b) with the selection button next to the menu item “More”.
In the service manual, the variant (a) is used.